7 Parameters
45
PRESSURE COMPENSATION
The MOD53 can accept a 4-20 mA input from a pressure transmitter to permit
automatic compensation for variations in the pressure. If it is available, set the “
P
Comp”
parameter to “
Auto”.
If not, set the
“P Comp”
parameter to “
Manual”
and
set the measurement pressure at the
“Input Pressure”
parameter. The units for
pressure display can be set to either “
Bar”
, “
Atm”, “psi”, “kPa”, “m H
2
O”
or “
mm
Hg”
at the
“P Units”
parameter. Variations of pressure can have a significant on
the sensor output..
SIMULATED INPUT
The facility exists within the MOD53 to simulate the input sensor levels to test the
set point and current output operation. This function allows the user to cycle up
and down through the sensor range using the “UP” and “DOWN” keys and display
the current output level, with the relays responding accordingly.
To select the
Simulate
menu item from the
Parameters
menu, press the “ENTER”
key. The unit will now display the simulation menu. Pressing the “UP” and
“DOWN” keys will cycle the displayed value between its minimum and maximum
levels in steps of 1%. The relays and current output will respond as if the displayed
value were an actual input, thus allowing the user to debug the set point and
current output configurations.
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...