9 Current Output
55
DUAL CURRENT OUTPUTS
If the instrument is supplied with two current outputs they are designated A & B
respectively. The menu for current output B is identical to A and is displayed after
the A menu.
PROPORTIONAL CONTROL
Many devices such as motor speed controllers, valve actuators, or stroke
positioners will accept an analogue 4-20 mA control signal.
It is possible to use the measurement signal from the instrument as a control
signal. By setting the point at which the output is 4 mA as the set point (e.g. 7.00
ppm) and the point at which the output is 20 mA as the proportional band (e.g. 4.00
ppm) a simple form of proportional control is achieved. If this signal was used to
drive a valve actuator, the valve would be fully open at 4.00 ppm; half open at 5.50
ppm and closed at 7.00 ppm.
ERROR CONDITION
The current outputs can be programmed to output 22mA or 0mA when an error is
detected on the source (i.e. Sensor Fault, Temperature Over or Under Range), to
provide remote warning of error conditions or to ensure fail safe operation. The
default state of is disabled, where the parameter is set to ”
No
”.
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...