9 Current Output
54
Ø
Ø
Ø
Ø
×
×
×
×
20.0
20.0
20.0
20.0
ppm
Current O/P Span
Span current (20mA) sensor input level
Ø
Ø
Ø
Ø
×
×
×
×
On Error =
No
22mA
0mA
Select state of current output when input
source error is detected (i.e. Sensor Input
Fault,
Temperature
input Over range)
INPUT
The input for the current output can be one of three options. The unit can use the
sensor input or the temperature input as the source for the zero and span.
Alternatively the output can be used to drive a cleaning system (see “Cleaning”).
When the cleaning option is selected, the span and zero menus disappear and are
replaced by the cleaning option menus.
OUTPUT RANGE
The output range for the current output can be set to one of two ranges, either 0 -
20mA or 4-20mA. This selection sets the limits of the zero and span output levels.
The output will continue to provide an extrapolated output above (>20mA) and
below (<4mA) these points but will flag an error message on the main display. The
maximum current limit is approximately 22mA, the minimum limit is 0mA (i.e. the
unit cannot source a negative current)
ZERO & SPAN
The zero and span levels are set as the limits of the source input. This provides a
totally flexible method of configuring the current output. The zero can be set
anywhere within the input source range and the span up to 5% of the selected
range, providing total control of the output range and offset. An inverse
relationship can easily be achieved by simply setting the zero level to be higher
than the span level.
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...