10 Calibration
63
CURRENT OUTPUT CALIBRATION
The user is provided with an opportunity to adjust the current output, to calibrate
any equipment that may be being used to monitor the current output signal. To
adjust the current output select the “
Current Output
” menu item in the
“
Calibration
” menu. Please keep in mind that the current output cannot go below
0mA. The maximum offset is
±
2mA. If two current outputs are fitted they are
referred to as A and B respectively.
Current Output
Current Output Calibration
(0,4 & 20mA levels)
Press to start calibration
Are You Sure?
The unit will request confirmation before
beginning the calibration
Press to confirm, any of the other 3 buttons
abandons the temperature calibration.
Adjust 0mA
Use the
Ø
and
×
keys to adjust the output to
give 0.00mA at the output monitor.
Press “ENTER” to finish.
Adjust 4mA
Use the
Ø
and
×
keys to adjust the output to
give 4.00mA at the output monitor.
Press “ENTER” to finish.
Adjust 20mA
Use the
Ø
and
×
keys to adjust the output to
give 20.00mA at the output monitor.
Press “ENTER” to finish.
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...