7 Parameters
43
P Units:
Atm
Bar
kPa
m H2O
psi
mm Hg
Pressure Display Units. Available when the
pressure compensation is set to “Manual”
Ø
Ø
Ø
Ø
×
×
×
×
1.0
1.0
1.0
1.0
Bar
Input Pressure
Fixed pressure level when
compensation is manual
Ø
Ø
Ø
Ø
×
×
×
×
Simulate Input
Press “ENTER” to Select the Input
Simulation Mode
1.0
1.0
1.0
1.0
ppm
Simulate : 12.00mA
Simulate sensor input. Use the “Up”
and “Down” keys to scan through the
input range to test the Set Point
Relays and Current Outputs
UNITS
The MOD53 primary display can be setup to display in % (saturation), ppm
(concentration), pO2 (partial pressure of Oxygen) or input current. This is achieved
by simply setting the appropriate units. The relationship between these three
parameters is determined by several factors including temperature, pressure and
the salinity of the solution being measured. (See Appendix A).
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...