10 Calibration
61
TEMPERATURE CALIBRATION
To initiate the temperature calibration select the “
Temperature
“ menu item in the
“
Calibration
” menu and press the “ENTER” key.
Temperature
Temperature offset calibration
(Only present if Compensation is Automatic)
Are You Sure?
The instrument will ask for confirmation
Press to confirm, any of the other 3 buttons
abandons the temperature calibration.
25.0
25.0
25.0
25.0
°C
The Temperature reading is now displayed.
Whilst the display flashes
Ø
Ø
Ø
Ø×
×
×
×
Adjust the reading using the up/down buttons
25.2
25.2
25.2
25.2
°C
Adjusted Reading
Press to accept the calibrated value when
ready.
Offset = +0.2°C
For user information, the unit will display the
offset adjustment
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...