Appendix I – Error Messages
103
Appendix I – Error Messages
Switch On Diagnostic Errors
E01 :
Internal Processor RAM Read/Write Error
Try switching the unit off then on again. If the message persists,
consult with your supplier, as this unit will require to be returned for
repair.
E02 :
EPROM Checksum Error
The software runs a checksum test on the program memory store at
power on, to ensure that the integrity of the software has not been
compromised. Switch the unit off then on again. If the error
message persists call LTH Electronics or authorised local
distributor.
E03 :
External RAM Read/Write Error
Try switching the unit off then on again. If the message persists,
consult with your supplier, as this unit will require to be returned for
repair.
E04 :
Setup Checksum Error
The instrument configuration has, for some reason, become
corrupted. Restore a setup from store A or B, or one of the two
default setups.
E05 :
Store A Checksum Error
The data in setup store A has been corrupted. Save the current
setup back to store A.
E06 :
Store B Checksum Error
The data in setup store B has been corrupted. Save the current
setup back to store B.
E07 :
Factory Calibration Checksum Error
The factory calibration data for this instrument has been corrupted.
The instrument will need to be re-calibrated. Consult your supplier.
E08 :
User Calibration Checksum Error
The user calibration data has been corrupted. Reset the user
calibration and re-enter
Summary of Contents for MOD53
Page 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Page 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Page 37: ...5 User Interface 36 This page intentionally blank...
Page 41: ...6 Main Display 40 This page intentionally blank...
Page 47: ...7 Parameters 46 This page intentionally blank...
Page 57: ...9 Current Output 56 This page intentionally blank...
Page 71: ...12 Pressure Input 70 This page intentionally blank...
Page 79: ...15 Fault Finding 78 This page intentionally blank...
Page 83: ...17 Guarantee and Service 82 This page intentionally blank...
Page 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Page 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Page 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Page 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Page 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Page 107: ...Appendix I Error Messages 106 This page intentionally blank...