43
SD 300_pH_1 01/2013
Set Corr: Adjustment of the measurements
**
Zero correction/offset of the voltage meas-
urement
**
oFF
No zero correction of the voltage
measurement
-10.0 - 10.0
mV
Zero correction of the voltage
measurement in mV
Slope adjustment of voltage measurement
**
oFF
No slope adjustment for voltage
measurement
- 5 . 0 0 0 -
5.000 %
Slope correction of voltage meas-
urement in %
Zero correction/offset of the temperature
measurement
**
oFF
No zero correction of the tempera-
ture measurement
-5.0 -
5.0%
Zero correction of the temperature
measurement in °C
Slope adjustment of temperature measure-
ment
**
oFF
No slope adjustment for tempera-
ture measurement
-5.00 -
5.00%
Slope correction of temperature
measurement in %
Set Alarm: Adjustment of the alarm function
On / No.So
Measurement channel pH/mV/rH:
Alarm on with sound / alarm on
without sound
OFF
No alarm function for measurement
channel pH/mV/rH
e.g.
- 2 . 0 0 -
14.00 pH
Min. alarm threshold pH/mV/rH
(not with AL. 1. oFF)
e.g.
-2.00 - 14.00
pH
Max. alarm threshold pH/mV/rH
(not with AL. 1. oFF)
On / No.So
Temperature measurement alarm
on with sound / alarm on without
sound
OFF
No alarm function for temperature
measurement
- 1 0 . 0 -
+110.0 °C
Min. alarm threshold temperature
(not with AL. 2. oFF)
- 1 0 . 0 -
+110.0 °C
Max. alarm threshold temperature
(not with AL. 2. oFF)