42
SD_300_pH_1 01/2013
Menu
Parameter
Values
Meaning
Auto Hold: Automatic measurement process-
ing (only effective with Logger = oFF mode)
on
Automatic measurement process-
ing (only with Logger = oFF) Auto
Hold
oFF
Standard hold function at the push
of a button (only with Logger = oFF)
Auto Power-Off : Automatic instrument
shut-off.
1 - 120
Shut-off delay in minutes. If no but-
ton is pressed and no data traffic is
taking place through the interface,
the instrument shuts off automati-
cally after the lapse of this time
oFF
Automatic shut-off deactivated
(continuous operation)
Background lighting
oFF:
No lighting
5 - 120
Automatic shut-off of lighting after
5 - 120 s (factory setting: 5 s)
on
Lighting always on
Universal output
SEr
Serial interface activated
dAC
Analogue output activated
oFF
Interface and analogue output off
-> minimal power consumption
01.11 - 91
Base address of the unit for serial
interface communication
e.g.
-2.00 - 14.00
pH
Input of the measurement at which
the analogue output should output
0V, e.g. with 0.00 pH
e.g.
-2.00 - 14.00
pH
Input of the measurement at which
the analogue output should output
1V, e.g. with 14.00 pH