41
SD 300_pH_1 01/2013
Menu
Parameter
Values
Meaning
rEAd Logg: Read of the individual logger data,
see Chapter 8!
Set Configuration: General settings
Input: Selection of the measurement
**
"rH" arrow
Measurement of the rH value
"mV" arrow Measurement of the mV value
(REDOX and/or ORP)
"mVH"
.
arrow
Measurement of the mV value
based on the hydrogen system
"pH" arrow
Measurement of the pH value
pH resolution: Resolution of the pH display
0.1 - 0.001
One tenth pH - one thousandth pH
Calibration: Selection of the number of cali-
bration points
1-Pt
1-point (only offset calibration,
incline -59.2 mV/pH)
2-Pt
2-point (n one additional
buffer)
3-Pt
3-point (n one one
additional buffer)
Calibration: Selection of the buffer series
Std
Standard buffer series (pH 7, pH 4,
pH 10) included in the "set" scope
of delivery
din
DIN 19266 buffer series pH 1.68(A),
pH 4.01(C), pH 6.87(D), pH 9.18(F),
pH 12.45(G)
Edit
Arbitrary buffer, manual setting
Calibration: Time interval for the calibration
memory (factory setting: 30 days)
1 - 365
Time interval for calibration mem-
ory (in days)
oFF
No calibration memory
Unit t: Selection of the temperature unit
**
°C:
All temperatures specified in de-
grees Celsius
°F:
All temperatures specified in de-
grees Fahrenheit