Figure 10. Electric connections
5.6. Electrical circuit of the monitoring unit
Electrical circuit of the monitoring unit is shown on the Figure 11.
The monitoring unit is purposed for measuring powering voltage and consumption current during maintenance procedure. It
also measures voltage of the tacho-generator and current of the motor at different speeds.
The monitoring unit consists of the terminal box and monitor.
The terminal box is transit unit with shunts Шн1 and Шн2 purposed for measuring consumption currents on them. Their
voltage drop is passed to the monitor through the connector Ш5.
The monitor has the measuring devices ИП1-ИП4 and switches for switching these devices. It has also 2-pin connectors for
external indicators (voltmeter, oscilloscope, frequency meter etc.)
The monitor has built-in marker generator assembled on the transistors T1-T4.
Powering of the marker generator is accomplished from the voltage stabilizer assembled on the resistor R7 and stabilitron Д2.
Switching the marker generator is accomplished by the B1 switch ‘Generator ON’.