: this message appears in the sample weight setting, in real calibration, after the eighth
sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press ESC to quit the
setting mode leaving the previous value unchanged. Examples: a number of decimals is
selected for full scale which exceeds the instrument's display potential; value above the
maximum setting value; the weight value set in sample weight verification does not
match the detected mV increase; the analog output correction goes beyond the permitted
limits.
:
lock active on menu item, keypad or display.
: It’s not possible to display properly the number because is greater than 999999 or less
than -999999.
: buffer battery low, loss of date and time of Real-Time Clock. Confirm with ENTER to
continue; leave the instrument on for at least 12 hours to charge the battery, if the alarm
persists contact technical assistance.
: an incorrect date has been detected: go into the related menu to check and correct it.
: waiting for START closure or ENTER key pressure to continue the batching.
:
PC is not connected.
: PC has not read batching data.
: gross weight equal to zero: the semi-automatic tare operation cannot be performed.