: the relay is not closed even in the presence of alarm
TEST
-
Input Test:
: ensure that for each open input
is displayed,
is displayed when the input is closed.
-
Output Test:
: setting
ensure that the corresponding output opens. Setting
ensure that the
corresponding output closes.
Refer to the following tables to activate the relay of various products (only 6 and 14 products
versions)
RELE6PROD MODULE
RELE14PROD MODULE
R1 R2 R3 R4 OUTPUTS
R1 R2 R3 R4 OUTPUTS
I
I
I
0 PROD.1
0
0
0
I PROD.7
0
I
I
0 PROD.2
I
0
0
I PROD.8
I
0
I
0 PROD.3
0
I
0
I PROD.9
0
0
I
0 PROD.4
I
I
0
I PROD.10
I
I
0
0 PROD.5
0
0
I
I PROD.11
0
I
0
0 PROD.6
I
0
I
I PROD.12
I
0
0
0 CYCLE END
0
I
I
I PROD.13
X
X
X
I SLOW
I
I
I
I PROD.14
-
E/EC Option Test:
: It shows the formula selected by the E/EC option, if the option is not present or is not
active, the message
is displayed.
-
Analog Output Option Test:
: It allows the analog signal to range between the minimum and the maximum values
starting from the minimum.
: current output test.
: voltage output test.
-
Millivolt Test:
: displays the load cell response signal in mV with four decimals.
+