Management and Recovery • 107
Gocator 2000 Family
Recovering Sensors
If a sensor’s network address or administrator password is forgotten, the sensor can be discovered
on the network and/or restored to factory defaults by using a special software tool called the Sensor
Discovery tool. This software tool can be downloaded from the downloads area of LMI’s website at
After downloading the tool package [14405-x.x.x.x_software_go2_tools.zip], upzip the file and run the
Sensor Discovery Tool [bin>win32>kDiscovery.exe]:
Any sensors that are discovered on the network will be displayed in the Devices List.
To change the network address of a sensor:
1
Select the sensor serial number in the Devices List.
2
Select the Custom Option.
3
Enter the new network address information.
4
Press the Set Address Button
To restore a sensor to factory defaults:
1
Select the sensor serial number in the Devices List
2
Press the Factory Restore... Button.
3
Confirm when prompted.
The Sensor Discovery tool uses UDP broadcast messages to reach sensors on different subnets. This enables
the Sensor Discovery tool to locate and reconfigure sensors even when the sensor IP address or subnet
configuration is unknown.
Summary of Contents for Gocator 2000 Family
Page 1: ...User s Manual Gocator 2000 Family Version 2 2 1 0 Revision A...
Page 164: ...Specifications 164 Gocator 2000 Family Envelope 61 25 40 5 4 15 82 9 95 7...
Page 166: ...Specifications 166 Gocator 2000 Family Envelope 33 90 80 2 9 84 3 99 7 10 6...
Page 168: ...Specifications 168 Gocator 2000 Family Envelope 190 210 33 141 3 155 7 4 11...
Page 170: ...Specifications 170 Gocator 2000 Family Envelope 33 300 400 214 9 229 2 10 3 3 2...
Page 172: ...Specifications 172 Gocator 2000 Family Envelope 33 400 500 216 2 230 6 3 9 11...