14
1.5 CombiLED with tap water connector
Type of problem
S
ca
ler
P
o
lis
h
e
r
Description
A. Unit is not responding and no lights are lit on the control
panel
X
X
Page 14
B. Lights are flashing on the control panel
X
X
Page 14-15
C. Difficulty to remove the ErgoGrip
X
X
D. Weak or no LED light
X
X
E. No tip vibration
X
F. Weak tip vibration
X
G. Insufficient or no water flow
X
H. Tip does not fit smoothly onto the handpiece
X
I.
Handpiece is overheated
X
J. Insufficient or no water flow
X
K. Insufficient or no powder flow
X
L. Water drips from the bottom of the device onto the table
X
X
A. Unit is not responding and no lights are lit on the control panel
1. Check that the power cord is connected properly and double check the voltage (120V/230V).
2. Check the fuse and replace if necessary. See Maintenance-section of the user guide.
3. Check that the wall outlet and the fuse panel of your clinic are OK.
4. If the problem still remains
– contact your dealer for support and indicate Error Code
B. Lights are flashing on the control panel
Five blue lights on the control panel are
flashing simultaneously
Check that the foot control cable is connected
at both ends and is not damaged.
The blue lights next to the S- and P-
keys are flashing alternately when
pressing the foot control
The Scaler/Polisher has not been switched
on. Press the S-key to select the scaler or
press the P-key to select the polisher.
The blue light next to the
S
-key is
flashing alternately with three other blue
lights on the control panel.
Check that the scaler handpiece is properly
connected.
The blue light next to the
P
-key is
flashing alternately with three other blue
lights on the control panel.
Check that the polisher tubing is properly
connected onto the connectors. Make sure it