10
1.3 AirLED with tap water connector
Type of problem
Description
A. Unit is not responding and no lights are lit on the control panel
Page 10
B. Lights are flashing on the control panel
Page 10
C. Difficulty to remove the ErgoGrip
D. Weak or no LED light
E. Insufficient or no water flow
F. Insufficient or no powder flow
G. Water drips from the bottom of the device onto the table
A. Unit is not responding and no lights are lit on the control panel
1. Check that the power cord is connected properly and double check the voltage (120V/230V).
2. Check the fuse and replace if necessary. See Maintenance-section of the user guide.
3. Check that the wall outlet and the fuse panel of your clinic are OK.
4. If the problem still remains
– contact your dealer for support and indicate Error Code
B. Lights are flashing on the control panel
Four blue lights on the control panel are
flashing simultaneously
Check that the foot control cable is connected
at both ends and is not damaged.
The blue light next to the ON-key is
flashing alternately with three other blue
lights on the control panel.
Check that the polisher tubing is properly
connected onto the connectors. Make sure it
is pressed in all the way.
Two yellow lights are flashing
alternately.
The Working Mode has not been selected.
Select the working mode by pressing one of
the keys next to the yellow lights.
The light next to the AP-key blinks twice
and pauses.
The device reminds of the need to activate
the cleaning cycle by pressing the AP-key (Air
Purge). Please see user manual for details.
The light next to the C-key is flashing
continuously.
The cleaning function has been activated.
Wait until the cleaning process has finished
and the light stops flashing.
C. Difficulty to remove the ErgoGrip from the handpiece
1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the
handpiece
2. If the problem still remains, replace the ErgoGrip