9
G. Difficulty to remove the ErgoGrip from the handpiece
1. Grab the ErgoGrip gently near the lens and at the same time twist and slide it off the
handpiece
2. If the problem still remains, replace the ErgoGrip
H. Handpiece is overheated during use
1. Check that the fluid flow is sufficient (at least 20 ml/min)
2. If the problem still remains, replace the handpiece and contact your dealer
I.
Weak or no LED light
1. Note that there are two versions of the ErgoGrip with different lenses and light pattern and
intensity: “ErgoGrip LED” and “ErgoGrip FocusLED”.
2. Check that lens in the ErgoGrip is clear and clean. Clean or replace if necessary.
3. If you have an additional handpiece, try replacing the handpiece
4. If the problem still remains
– contact your dealer for support and indicate Error Code