10.4. Operational test
Additional equipment needed: Artificial denture, cup for water flow, timer.
Task
Description
Pass/Fail
Initials
Date
Connect cable to
mains
Green standby light is illuminated
Select mode
Select scaler mode by pressing S on control
panel.
Test Scaler Irrigation
mode
Test the water regulation of the handpiece by
selecting irrigation mode and testing handpiece
water flow ring positions. Continuous mist
and dripping should be reached according the
control ring angle.
Test scaling with
water -mode
Select WM1 (40%) and test foot pedal
response. Use artificial denture in this test.
Test scaling with
water -mode
Select WM2 (70%) and test foot pedal
response. Use artificial denture in this test.
Test scaling with
water -mode
Select WM3 (100%) and test foot pedal
response. Use artificial denture in this test.
Test dry mode
Select 40% mode and test the foot pedal
response. Use artificial denture in this test
Test dry mode
Select 70% mode and test the foot pedal
response. Use artificial denture in this test.
Test dry mode
Select 100% mode and test the foot pedal
response. Use artificial denture in this test.
Test Polisher
Select working mode 2 and test polishing on
the artificial denture.
Select working mode 3 and test polishing on
the artificial denture.
Test diagnostic
function
Set LED on and off with the foot pedal.
Test Air-Blow mode
Select air-blow mode by pressing P. Test air
blow on the artificial denture.
Test Air Purge
function
Select polisher mode by pressing P on control
panel. Press C on control panel and record
function.
After all the above mentioned tests are passed the instrument is ready for the routine operation.