Cleaning and maintenance
32
Autoclave in steam at 134°C (max 135°C) for minimum 3 minutes
Scaler tips
Polisher nozzle
Torque wrench
LM-ErgoGrip Focus LED
LM-ProPower Scaler handpiece
7.4. Recommended cleaning procedure
Beginning of the day
Run the automatic cleaning cycle with clean water for both scaler and polisher. See instructions chapter “7.2.
Automatic cleaning function for the scaler and polisher” on page 30.
After each treatment
• To prevent clogging, air purge the polisher handpiece after each treatment. See instructions “6.4. Air
purge quick cleaning function” on page 27.
• Run the automatic cleaning cycle with clean water for both the scaler and polisher. See instructions “7.2.
Automatic cleaning function for the scaler and polisher” on page 30.
• Clean off the cover, control panel, handpieces and the hoses with a soft cloth. Use a surface disinfectant
suitable for hard plastics.
• Clean the nozzle in an ultrasonic bath (40-50°C), for minimum 3 minutes, before cleaning/sterilization. If
ultrasonic cleaning is not possible, flush the nozzle with warm water.
• Clean the ErgoGrip, the tip, the nozzle and possibly the scaler handpiece and autoclave according to the
clinic’s routines.