![LM-Dental LM-ProPower CombiLED User Manual Download Page 26](http://html1.mh-extra.com/html/lm-dental/lm-propower-combiled/lm-propower-combiled_user-manual_1932118026.webp)
Operating instructions AirLED Polisher
26
AirLED Polisher
A
1. Unscrew the powder container from the cap and fill
the powder container with LM-ProPower powder
until MAX (A).
NOTE: With LM-Glycine powder fill the powder
container only half of the maximum to prevent too
much powder consumption and dispersion.
2. Screw the container back in to the cap.
3. Connect the powder container cap to the
connectors on the right side of the unit (B).
4. Bottle version: Connect the medicament bottle to
the unit.
5. Carefully move the LM-ErgoGrip onto the polisher
handpiece (C).
6. Screw the polisher nozzle clockwise into the
handpiece until you feel a stop (D).
7. Press the key identified P on the control panel (E) to
Set the polisher ON. The blue polisher indicator lamp
illuminates.
8. The indicators adjacent to the working modes
keys are flashing to remind that you must select
a working mode. If some blue indicators are
flashing alternatively, make sure that the polisher is
connected. If the problem stays, refer to the trouble
shooting section. Choose working mode 2 or 3
by pressing the related working mode key on the
keyboard.
9. Adjust water flow for optimal performance.
• Point the handpiece to the cuspidor and above the
bowl. Push the foot switch to activate the polisher
and adjust the water flow on the powder container
(F).
• Hold the polisher nozzle approximately 1 cm (0,4
in.) down right from the bottom of the bowl and
press the foot control to activate the polisher.
• Slowly reduce the water flow until the powder starts to
collect on the surface as a white spot.
• Increase the water flow until the spot just
disappears. The air polisher is now balanced for
optimum performance.
B
C
D
E
F