19
USER GUIDE
· Do not touch the batteries during measurement.
· This device is portable, but should not be operated in transit.
· You can use this device to take your own measurement, no third-party operator is required.
· In rare occasions, the device may remain inflated after a measurement. If this were to occur, make sure
to disconnect the tube from the device to release pressure. Prolonged pressurization of the cuff may
lead to ecchymosis or other injury.
· Do not kink, fold, stretch, compress, or otherwise deform the tube during measurement as it may lead
to an inaccurate result, or excessive inflation of the cuff which may cause injury.
· Accuracy of pressure measurement (at 50 and 200 mmHg) and the rate of air leakage from the system
should be checked after repair, maintenance, and every two years of use.
· The device must only be serviced and repaired by authorized service personnel.
· At the request of authorized service personnel, circuit diagrams, component part lists, descriptions,
and calibration procedures will be made available by the manufacturer or distributor.
· Do not use the device while under maintenance, or being serviced.
· Dispose of the device, its accessories, and batteries according to your local waste disposal
regulations and guidelines.