· The device may require up to 30 minutes to warm up/cool down from the minimum/maximum
storage temperature before it is ready for use.
· Do not use the device where flammable gases are present, or in environments with high
oxygen concentration.
· Measurements may not be accurate if the device is used in close proximity to devices generating strong
electromagnetic fields, such as microwave ovens, wireless charging stations, X-ray machines, etc.
· Wireless communication devices, such as Wi-Fi access points, mobile phones, and cordless phones may
cause interference that may affect the accuracy of measurements. A minimum distance of 1 ft (30 cm)
should be kept from such devices during measurement.
· The device is not intended for use on extremities other than the arm, or for any purpose other than
measuring blood pressure and pulse rate.
· Do not confuse self-monitoring with self-diagnosis.
· This device allows you to monitor blood pressure.
· Do not begin or alter medical treatment without consulting a medical professional.
· Do not take frequent, consecutive measurements as these may constrict blood flow and lead to injury.
· This unit is not suitable for continuous monitoring during medical emergencies or operations.
· This device is not anesthesia proofed (AP/APG), and it is not suitable for use in the presence of
flammable anesthetics.