D9U001SFS-0101_03
34
9.2.1 Instructions for putting into service
►
Place the Hub Holder on the accessory rail on the right side of the Image 3 bed.
►
Tighten the safety bolt to fix Hub Holder on the accessory rail. Check if the Hub Holder is safely installed on the
accessory rail.
►
Bolt the Holder of the interconnecting elements for Nurse Call cable and LAN cable to the crossbar at head end of
the Image 3 bed.
►
Insert both interconnecting elements for Nurse Call cable and LAN cable to the Holder of the interconnecting
elements in any order.
►
Connect LAN cable, Nurse Call cable and Wetness Pad cable to the back side of the Hub box.
►
Tighten grommets of the connected cables on the back side of the Hub box by turning.
►
Secure the connected cables by inserting comb-like part.
►
Place the Hub to the Hub Holder and ensure the Hub is fixed in the Hub Holder by the raised latch.
►
Disconnect the Power Supply/Data cable from the safety position behind the accessory rail.
►
Remove the plug covering the connector on the right side of the Hub.
►
Connect Power Supply/Data cable to the connector on the right side of the Hub.
►
Route LAN cable and Nurse Call cable above the accessory rail towards the head end of the bed. Place both ca
-
bles to the plastic holders on the frame of the mattress support platform.
►
Connect LAN cable and Nurse Call cable to the interconnecting elements for Nurse Call cable and LAN cable at
head end of the Image 3 bed according to the size of the connectors.
►
Raise the Backrest of the Image 3 bed to the highest position.
►
Fix the Air Pad on the Backrest so that the 2 middle plastic sections of the Backrest are surrounded by the straps
of the Air Pad Cover.
►
Connect the Air Hose to the right side of the Hub.
►
Route the Air Hose along the two cables above the accessory rail towards the head end of the bed.
►
Run the Air Hose from below up through the hole in the hook on the accessory rail. The hook selected for this
purpose must be put on the accessory rail in the other way around so it is not possible to use this hook for carrying any
accessory. The position of the selected hook must be adjusted with respect to the Air Pad position on the Backrest in the
highest position.
►
Run the Air Hose upwards through the plastic sections of the Backrest.
►
Connect the Air Hose to the Air Pad fixed on the Backrest.
►
Ensure the Backrest adjustment does not damage the Air Hose and the Air Hose is not squeezed in any position
of the Backrest.
►
Place a compatible mattress on the mattress support platform of the Image 3 bed.
►
Place Wetness Pad on the mattress without any sheet to the correct position in the seat section of the patient.
Four connectors on the Wetness Pad are facing a surface of the compatible mattress.
►
Connect Wetness Pad cable to the Wetness Pad according to the four labels on the Wetness Pad cable. There are
two ways of connection depending on the starting position of the Wetness Pad that are in accordance with the correct
use.
Slip the overlapping ends of the Wetness Pad on the both sides of the Image 3 bed under the mattress.
►
Place a sheet on the mattress.
►
Connect the cable connected to the local Nurse Call system to the interconnecting element for Nurse Call cable at
head end of the Image 3 bed, if the use of the Nurse Call system with SafeSense® 3 system is required.
►
Connect the local LAN cable to the interconnecting element for LAN cable at head end of the Image 3 bed, if wire
-
less signal from Hub is not used for communication between Hub and a hardware with installed CareMonitor Application.
►
Place patient on the Image 3 bed.
►
Activate Hub according to the instructions for use.
►
Open the CareMonitor Application.
►
Activate the Hub and check if the corresponding bed is visible on the CareMonitor Screen.
►
Perform the same steps to activate the other Hubs and to display the corresponding beds on the CareMonitor
Screen.