9.3.1 Instructions for putting into service
►
Place the Hub Holder on the accessory rail on the right side of the Eleganza 2 bed so that the holder with interco
-
nnecting elements for Nurse Call cable and LAN cable will be accessible next to the left side of the Hub Holder.
►
Tighten the safety bolt to fix Hub Holder on the accessory rail. Check if the Hub Holder is safely installed on the
accessory rail.
►
Connect LAN cable, Nurse Call cable and Wetness Pad cable to the back side of the Hub box.
►
Tighten grommets of the connected cables on the back side of the Hub box by turning.
►
Secure the connected cables by inserting comb-like part.
►
Place the Hub to the Hub Holder and ensure the Hub is fixed in the Hub Holder by the raised latch.
►
Disconnect the Power Supply/Data cable from the safety position behind the accessory rail.
►
Remove the plug covering the connector on the right side of the Hub.
►
Connect Power Supply/Data cable to the connector on the right side of the Hub.
►
Connect LAN cable and Nurse Call cable to the interconnecting elements for LAN cable and Nurse Call cable in
the holder next to the left side of the Hub. LAN cable must be connected to the left white socket. Nurse Call cable must be
connected to the right black socket.
►
Raise the Backrest of the Eleganza 2 bed to the highest position.
►
Fix the Air Pad on the Backrest so that the Air Pad will be placed over the space between the 2 plastic parts on
the Backrest and the fixing straps will be led through the 4 holes in the 2 different plastic parts.
►
Connect the Air Hose to the right side of the Hub.
►
Route the Air Hose above the accessory rail towards the head end of the bed.
►
Run the Air Hose through the hole in the left accessory rail holder next to the holder with interconnecting ele
-
ments for Nurse Call cable and LAN cable.
►
Run the Air Hose upwards and then back above the accessory rail in order to run the Air Hose between the Bac
-
krest and the Seat section.
►
Run the Air Hose towards the head end of the bed in groove of the Backrest plastic part.
►
Connect the Air Hose to the Air Pad fixed on the Backrest.
►
Ensure the Backrest adjustment does not damage the Air Hose and the Air Hose is not squeezed in any position
of the Backrest.
►
Place a compatible mattress on the mattress support platform of the Eleganza 2 bed.
►
Place Wetness Pad on the mattress without any sheet to the correct position in the seat section of the patient.
Four connectors on the Wetness Pad are facing a surface of the compatible mattress.
►
Connect Wetness Pad cable to the Wetness Pad according to the four labels on the Wetness Pad cable. There are
two ways of connection depending on the starting position of the Wetness Pad that are in accordance with the correct
use.
Slip the overlapping ends of the Wetness Pad on the both sides of the Eleganza 2 bed under the mattress.
►
Place a sheet on the mattress.
►
Connect the cable connected to the local Nurse Call system to the interconnecting element for Nurse Call cable at
head end of the Eleganza 2 bed, if the use of the Nurse Call system with SafeSense® 3 system is required.
►
Connect the local LAN cable to the interconnecting element for LAN cable at head end of the Eleganza 2 bed, if
wireless signal from Hub is not used for communication between Hub and a hardware with installed CareMonitor Applica-
tion.
►
Place patient on the Eleganza 2 bed.
►
Activate Hub according to the instructions for use.
►
Open the CareMonitor Application.
►
Activate the Hub and check if the corresponding bed is visible on the CareMonitor Screen.
►
Perform the same steps to activate the other Hubs and to display the corresponding beds on the CareMonitor
Screen.