D9U001SFS-0101_03
68
Indicator
Meaning
Warning will be triggered on each CareMonitor
user interface connected to the SafeSense® 3
system when corresponding status occurs during
activated monitoring (WARNINGS TURNED ON).
Warning will
not
be triggered on any CareMonitor
user interface connected to the SafeSense® 3
system when corresponding status occurs during
activated monitoring (WARNINGS TURNED OFF).
Warnings visible on a single CareMonitor user in-
terface and audible from a single CareMonitor user
interface and sent to Nurse Call system.
Warnings not visible on a single CareMonitor user
interface and not audible from a single CareMo-
nitor user interface, but sent to the Nurse Call
system.
Click on this button to zero the countdown of the
Decubitus/Movement Notifi cation. If a set period
is changed, the countdown related to the previous
setting is not reset. RESET button serves to reset
the running countdown.
Abbreviation
Meaning
Wetness
Wetness Monitoring
Motion
Motion Monitoring
Decubitus
Decubitus/Movement
Notifi cation
Bed Exit
Bed Exit Monitoring
RANGES OF VALUES
Periods for Bed Exit Monitoring are
early
(about
3
seconds) or
10
seconds or
20
seconds or
30
seconds or
40
seconds or
50
seconds or
1
minute or
5
minutes or
10
minutes or
15
minutes or
30
minutes.
Periods for Motion Monitoring are multiples of
5
minutes from
10
minutes
to
100
minutes.
Periods for Decubitus/Movement Notification are
2
hours or
3
hours or
4
hours or
5
hours or
6
hours.
Limits of the wetness sensitivity are
25%
or
50%
or
75%
or
100%.
GLOBAL WARNINGS SETTINGS
To set Sensitivity of the Bed Exit Monitoring and Motion Monitoring:
► Click on
low
button or
medium
button or
high
button.
Fig. Global warnings settings