D9U001SFS-0101_03
84
SafeSense® 3 Display (main screen)
8
7
1
1
2
2
5
4
4
3
3
7
6
6
5
Fig. SafeSense® 3 Display - monitoring ON
Fig. SafeSense® 3 Display - monitoring OFF
1.
Activation Symbol (selected type (or types) of
monitoring is (are) activated)
2.
Indicator of communication between Hub and
Attendant Control Panel (black arrows - communication
works, orange cross over black arrows - communication
does not work)
3.
Indicator of discharged Hub battery
4.
Indicator of activated Motion Monitoring
5.
Period during which the absence of patient
movement on the bed is tolerated. Countdown to the
selected time (e.g. 4 hours) starts again from the begi-
nning after patient movement on the mattress.
6.
Period during which the Bed Exit alerts are
not triggered unless patient is on the bed (e.g. no alert
during 30 minutes)
7.
Indicator of activated Bed Exit Monitoring (pati
-
ent presence on the mattress is monitored)
8.
Indicator of Wetness Monitoring (Wetness Mo-
nitoring is activated)
1.
Deactivation Symbol
2.
Indicator of communication between Hub and
Attendant Control Panel (black arrows - communication
works, orange cross over black arrows - communication
does not work)
3.
Indicator of activated Decubitus/Movement
Notification
4.
Period after which nursing staff is notified to
check position of the patient
5.
Period during which the Bed Exit alerts are
not triggered unless patient is on the bed (e.g. no alert
during 30 minutes)
6.
Indicator of activated Bed Exit Monitoring (pati
-
ent presence on the mattress is monitored)
7.
Indicator of Wetness Monitoring (Wetness Mo-
nitoring is activated - The more black symbol of drop,
the higher degree of the wetness is monitored.)
Orange square symbol is displayed instead of the green activation symbol when no monitoring is activated.
Indicator of Wetness Monitoring is not displayed when Wetness Monitoring is not activated.
Indicator of Wetness Monitoring changes according to the degree of monitored wetness. The more black symbol of drop, the higher
degree of the wetness is monitored.