4.Advanced User Guidebook
66
menu as
Start
, the test will begin.
6.
Save rule:
Select
SaveRule
in the bottom menu. Select the save location in the
left menu, and then select
Save
in the right menu to save the rules, which could be
called up at once when need. Select
Load
to call up the rule saved.
Note:
1.
When Pass/Fail is ON, if XY or FFT is ready to run, then Pass/Fail will be closed;
under the mode of XY or FFT, Pass/Fail is unable.
2.
Under the mode of Factory, Auto Scale and Auto Set, Pass/Fail will be closed.
3.
When no save setting left in the rule save, tip will be given to show "NO RULE
SAVED".
4.
Under the status of stop, data comparing will stop, and when it goes on running,
the number of Pass/Fail will increase from the former number, not from zero.
5.
When the waveform playback mode is on, Pass/Fail is used to test the
played-back waveform specially.
●
Output
Push the
Utility
button, select
Function
in the bottom menu, select
Output
in the left
menu.
Output
menu item in the bottom menu sets the output type of Trig Out(P/F) port in
Figure 3-3 Rear Panel on P6. In the bottom menu, select
Output
. The description of
Output menu
is shown as the follows:
Video
menu item in the bottom menu sets the output port of video. In the bottom
menu, select
Video
. The description of
Video menu
is shown as the follows:
Device
and
Print Setup
menu items set the print output, refer to "
●
LAN Set
Using the LAN port, the oscilloscope can be connected with a computer. Refer to
"
" on page 80 for the operation steps.
Function
Menu
Setting
Description
Type
Trig level
Output trig signal synchronously
Pass/fail
Output High Level when Pass , and Low Level when Fail
Function
Menu
Setting
Description
Video
OFF
VGA
AV
Turn off video output
Connect the VGA or AV port to an external monitor or
projector. Select the port in this menu, the oscilloscope
display can be shown on an external monitor or projector.