4.Advanced User Guidebook
33
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
knob to set time interval before
another trigger occur, press
panel button to
move cursor to choose which digit to be set.
Set Holdoff time as 100 ns
7.Timeout Trigger
The oscilloscope triggers when the time interval from when the rising edge (or the
falling edge) passes through the trigger level to when the neighbouring falling edge
(or the rising edge) passes through the trigger level is greater than the timeout time
set.
In Timeout Trigger mode, the trigger setting information is displayed on bottom right
of the screen, for example,
,
indicates that trigger type is Timeout,
trigger source is CH1, edge is positive, 0.00mV is up level or low level threshold.
Timeout Trigger menu
list:
MENU
SETTING
INSTRUCTION
Single
Mode
Timeout
Set vertical channel trigger type as Timeout trigger.
Source
CH1
CH2
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Edge
Edge
Start timing when the rising edge of the input signal
passes through the trigger level.
Start timing when the falling edge of the input signal
passes through the trigger level.
Configure
Idle Time
Set idle time. Idle time means the minimum time of idle
clock before searching data that can meet trigger
conditions. Available range is 30ns-10s, default 100ns.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
knob to set time interval before
another trigger occur, press
panel button
move cursor to choose which digit to be set.
Set Holdoff time as 100 ns
8.Nth Edge trigger
The oscilloscope triggers on the Nth edge that appears on the specified idle time. As
figure shown below, the oscilloscope should trigger on the second falling edge after