53
VIII. Technical Specifications
i. Set the probe compensation at x10 for physical probe;
ii. Under suitable operation temperature, work the device for 30+ minutes without interval;
iii. Provided the environment temperature gets 5 degrees more, or less, to self-calibrate the device
through Utility menu, via “Self Cal” option (please refer to
).
The following technical specifications been measured on the basis of above operation,
Oscilloscope Part -
Bandwidth
VDS6102
100MHz
VDS6152
150MHz
VDS6102A
8-bit mode
100 MHz
12-bit mode
100 MHz
14-bit mode
20 MHz
VDS6152A
8-bit mode
150 MHz
12-bit mode
100 MHz
14-bit mode
20 MHz
Vertical Resolution (A/D)
VDS6102
VDS6152
8 bits
VDS6102A
VDS6152A
8 bits / 12 bits / 14 bits
Channel Q'nty
2
Acquisition
Mode
sample, peak detect, average
Sampling Rate
VDS6102
VDS6152
2-CH working
500 MS/s
1-CH working
1 GS/s
VDS6102A
VDS6152A
2-CH working
8-bit mode
500 MS/s
12-bit mode
250 MS/s
14-bit mode
100 MS/s
1-CH working
8-bit mode
1 GS/s
12-bit mode
500 MS/s
14-bit mode
100 MS/s
Input
Input Coupling
DC, AC, ground
Input Impedance
1 MΩ ± 2%, in parallel with 15 pF ± 5 pF
Supported Probe
x1, x10, x100, x1000
Max Input Voltage
40V (DC + AC Peak)
Bandwidth Limit
20MHz, or fullband
Channel Isolation
100 : 1 @ 50Hz; 40 : 1 @ 10MHz
Time Delay Between
Channel (typical)
150 ps