Lighthouse REMOTE Active Count Operating Manual
A-6
248083401-1 Rev 5
associated with location index values. Thus the name for location =3
would be located at registers 40208-40211. Up to two hundred
locations can be specified.
Register 40199 indicates the number of location names supported on
this device.
Device Status
The Device Status register (40003) displays the current status of the
device.
Command
Register
The Command Register (40002) is used to make the device perform an
action. The register performs an action when an integer value is written
to it. The action is completed when the device sends a MODBUS
response. When this register is read, it always returns a zero.
Table A-6 Device Status
Bit
Description
0
RUNNING:
Set when a start command is executed
remotely via Command 9 (manual start) or Command 11
(instrument start) or through the user interface. The flag
will remained set until a stop command is executed.
1
SAMPLING:
This is set only when the instrument is
actually sampling data that is to be recorded. Caution
must be used in sending a command during this time that
may invalidate current sample.
2
NEW DATA:
Set to 1 to indicate that a new data record
has been recorded and it hasn't been read via modbus yet.
When a data record has been read via modbus (registers
30001 to 30999), then this flag is reset to zero.
12
FLOW STATUS: Set to 1 when unit’s flow is out of spec,
else 0.
Table A-7 Command Register
Value
Action
1
Saves all writable 4xxxx register values to the EEPROM.
2
Reserved for future use.
3
Clears the Data Buffer. Record count is set to zero.
Summary of Contents for REMOTE Active Count
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Page 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Page 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Page 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Page 61: ...00...
Page 62: ......