Lighthouse REMOTE Active Count Installation Guide
3-12
248083401-1 Rev 5
Configure the flow settings as explained in steps 9 through 13. As
each step is completed, the Save function will cause the LED to
blink again.
Figure 3-16 Flow Adjust Controls
8. Before making any changes to settings, watch the flow meter and
press down firmly on the Flow Adjust Cover to determine if any
fluctuations in the flow occur. If fluctuations do occur, rotate the
Cover, press down and monitor the flow meter until a steady state is
acheived. Rotate the Impactor top piece and Cover until the highest
steady state is acheived. Use an automatic center punch to place a
mark on the outside rim of the top piece and the base, adjacent to
each other.
9. If no fluctuations occur, turn the Impactor top piece to determine if
any fluctuations occur. Turn the top piece until the highest steady
vacuum is achieved then use an automatic center punch to punch
two marks, one on top piece and other on bottom piece adjacent to
each other.
Note:
The center punch
marks provide alignment
markers for mating the
pieces after cleaning or
dismantling.
10. When the LED stops blinking, the median (28.3 lpm) flow setting is
ready to be set. Use the flow meter attached to the Impactor and
adjust the flow knob on the bottom of the RAC until the meter reads
28.3 lpm (1 CFM) and is steady. Briefly press the Start/Stop button
to save, which blinks the LED again.
11. When the LED stops blinking, the upper flow setting can be set.
Adjust the flow knob until the meter reads 31.1 lpm (1.1 CFM) and
is steady. Briefly press the Start/Stop button to save, which blinks
the LED briefly.
12. When the LED stops blinking, the lower flow setting can be set.
Adjust the flow knob until the meter reads 25.5 lpm (0.9 CFM) and
is steady. Briefly press the Start/Stop button to save the parameters.
1 - Press & Hold Switch
2 - Turn Rocker Switch ON
3 - Flow Setting Knob
Summary of Contents for REMOTE Active Count
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Page 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Page 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Page 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Page 61: ...00...
Page 62: ......