MODBUS Register Map v1.49
248083401-1 Rev 5
A-5
Alarm Mode (40037) defines the type of calculation performed to
define an alarm condition. Alarm Mode = 0 corresponds to
conventional threshold alarming; channel bit set if threshold exceeded
for that given channel.
Alarm Parameter (40038) defines additional parameters that may be
needed in defining an alarm mode.
Printer Options (40049)
displays the configuration of the instrument’s
printer function.
If Bit-1 of Register 40049 is set, the instrument will print the last
recorded data at the end of each sample. This feature cannot be enabled
if the
One Second Data Update
feature is enabled.
Device Options (40050)
displays the instrument’s device configuration.
If bit 0 of Register 40050 is set, it indicates that the instrument is
capable of Fast Download.
If Bit 1 of Register 40050 is set, the instrument will display and update
the data registers every second. No data will be recorded in the data
buffer. Enabling this feature disables the Print on Sample feature.
Registers 40200-40999 are reserved for eight character names
40999
ASCII string
Location_200_char[6], char[7]
Table A-3 Sensor Settings Registers
Register Data
Type
Description
Table A-4 Printer Options
Bit
Description
0
Unused - non-writeable
1
Print on Sample (1=Enabled, 0=Disabled)
2-15
Reserved
Table A-5 Device Options
Bit
Description
0
Fast Download (1=Enabled, 0=Disabled) non-writeable
1
One Second Data Update (1=Enabled, 0=Disabled)
2-15
Reserved
Summary of Contents for REMOTE Active Count
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Page 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Page 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Page 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Page 61: ...00...
Page 62: ......