Installation
Basic hand and power tools are required to install the instrument in a
permanent location. It should be mounted on a flat vertical surface and
the Inlet-to-Outlet path must be as vertical as possible.
Choose a location that is convenient to all of the facilities required,
such as electricity, vacuum and network connections and provides six-
inch clearance on the left and right sides. Use of safety glasses and
gloves are strongly suggested.
WARNING:
REMOTE 1100/1104 Power Supply specifications
are 100-240VAC, 47-63Hz, 1.4A input, 24VDC 3.0A output
(72W max). Installation of the REMOTE must be done with
easy access to the DC power connector, enabling fast
manual DC removal from the REMOTE. The power supply
must be installed such that the AC power cord is not
obstructed and AC power can be quickly removed.
If replacement of the power supply or its AC power cord is
required, replace only with a power supply or cord having
as good as or better ratings than the items provided by
Lighthouse Worldwide Solutions. Attempting to use an
under-rated power supply or cord can expose the
instrument power supply, adjacent equipment or the user to
dangerous shock and fire hazards. Failure to heed this
warning can result in personal injury or death.
Every installation of measuring instruments is dependent upon facility
surroundings, construction materials and points of placement. Due to
these differences, specific instructions cannot be supplied in this
document. The listed tools and hardware shown are typical but may not
be used for every situation.
Figure 3-1 Required Tools
Maintain a minimum
clearance of six
inches from the left
and right sides of
the instrument to
prevent overheating
of the instrument
and subsequent
failure. Failure to
heed this warning
may result in the
warranty being void
and any repairs will
be charged to the
customer.
Bubble Level
Screw Driver
Pencil or pen
Instrument as a Template
Wall Anchor Devices
Items not shown:
Battery-powered
Drill Motor
Summary of Contents for REMOTE 1100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Page 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Page 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Page 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Page 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Page 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Page 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Page 69: ...00...
Page 70: ......