Programming
248083363-1 Rev 4
4-7
Running the
Instrument
Action commands for running the
REMOTE
are discussed here:
There are 2 basic modes of operation: Manual counting and Automatic
counting. Write each of the described action commands to the
command register (40002).
Table 4-6 Action Commands
Value
Action
1
Saves all writable 4xxxx register values to the
EEPROM.
3
Clears the Data Buffer. Record count is set to zero.
4
Saves the instrument parameters in the 40xxx registers
to the EEPROM. Parameters include Sample Time,
Hold Time, Initial Delay, and Location.
5
Enable Remote Control. Locks out the instrument’s
user interface. Can only change instrument
parameters via Modbus.
6
Enable Local Control. Unlocks the instrument’s user
interface. Instrument changes can be made at the
device itself or through Modbus.
7
Start pump.
8
Stop pump.
9
External Start Counter. The instrument samples
continuously until it receives an External Stop Counter
command. Does not turn on the pump. Ignores local
timing parameters.
10
External Stop Counter. Records counts since External
Start.
11
Instrument Start. Uses defined Initial Delay, Hold
Time, Sample Interval and counting mode. Instrument
executes samples and holds until an Instrument Stop
command is issued. This command will start the
pump.
12
Instrument Stop. Aborts current sample. Stops pump.
Stops data collection.
Summary of Contents for REMOTE 1100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Page 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Page 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Page 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Page 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Page 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Page 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Page 69: ...00...
Page 70: ......