Introduction
248083363-1 Rev 4
2-3
REMOTE 1100
Specifications
Table 2-1
REMOTE 1100
Specifications
Size Range
0.1 - 1.0
μ
m
Channel Thresholds
Standard
: 0.1, 0.2, 0.3, 0.5
μ
m
Flow Rate
1.0 CFM (28.3 LPM)
Counting Efficiency
50% (per JIS)
Light Source
Long-Life Laser Diode
Zero Count Level
<1 count/5 minutes (per JIS)
Vacuum Source
Customer-supplied Vacuum >18”Hg
Calibration
NIST Traceable, per ISO 21501-4
Communication
Mode
CEMS (pulse output)
Data Storage
999 records, rotating buffer
Supporting Software
Lighthouse Monitoring System, LMS XChange
Power Supply
50-60Hz, 100-240VAC @1.5A; output 24VDC @ 3.0A
(72W); must be LPS and certified to CE requirement
Power Input
24 [email protected] min (38W min) and 2.5A max
(60W)
Enclosure
Stainless Steel
Dimensions
10.5"[w] x 7.65"(h) x 3.35"(d) [26.67 x 19.43 x 8.51
cm]
Weight
7.0 lbs (3.18 kg)
Operating Temp/RH
50° F to 104° F (10° C to 40° C) / 20% to 95% non-
condensing
Operating Altitude
Up to 2000m
Storage Temp/RH
14° F to 122° F (-10° C to 50° C) / Up to 98% non-
condensing
Summary of Contents for REMOTE 1100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Page 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Page 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Page 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Page 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Page 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Page 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Page 69: ...00...
Page 70: ......