Lighthouse REMOTE 1100/1104 Operating Manual
3-4
248083363-1 Rev 4
6. Connect the supply to AC and observe the REMOTE LEDs. Power
should come on and stay on; Flow should blink then go off. Service
should turn on solid.
7. Connect to communication network.
8. Remove Inlet and Outlet caps, connect vacuum source and purge
filter then verify correct Flow condition after a couple of minutes.
Allow flow for several hours to purge instrument.
Operation
Before relocating the counter, power it OFF and reinstall the caps to
prevent contamination of the LASER sensor.
Understanding the LEDs
The front-panel LEDs have special meanings when illuminated. The
figure below shows locations of the LEDs and a description of their
meaning.
Figure 3-1 Front Panel LEDs
•
The green POWER LED turns on when the instrument is powered
on.
•
The green FLOW LED turns ON when the flow is correct, BLINKs
when flow is insufficient or is zero.
Summary of Contents for REMOTE 1100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Page 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Page 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Page 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Page 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Page 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Page 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Page 69: ...00...
Page 70: ......