Lighthouse REMOTE 1100/1104 Operating Manual
4-2
248083363-1 Rev 4
DIP Switch
Settings
GENERAL DEFINITIONS
OFF (LEFT) = 0, ON (RIGHT) = 1
Note:
The DIP Switches
must be set before the unit
is powered on.
Communications Mode (DIP #6)
In order to communicate with the instrument using the Modbus
protocol, DIP switch 6 must be set to ON.
Addressing (DIP 1-5)
Note:
Because Address
’0’ is reserved for RS-485
broadcast. Whenever all
of the dip switches are
OFF or when DIP switch1
is ON, the instrument
address is set to ’1’.
Table 4-2 details the addresses set by the DIP switches 1-5.
Table 4-1 DIP Switch settings
Position#
Description
Setting
1
Binary Bit 0
OFF=0, ON=1
2
Binary Bit 1
OFF=0, ON=1
3
Binary Bit 2
OFF=0, ON=1
4
Binary Bit 3
OFF=0, ON=1
5
Binary Bit 4
OFF=0, ON=1
6
Communication
Mode
OFF: Pulse Mode
ON: Modbus Mode
7
Count Mode
Pulse Mode: OFF: CUML Count
ON: DIFF Count
Modbus Mode: Reserved
8
Fast Transfer
Mode
(Save to Buffer?)
Pulse Mode: Reserved
Modbus Mode: OFF: Data Saved
ON: No Data Saved
Table 4-2 DIP Switch Addressing
DIP SWITCHES
1 2 3 4 5
ADDRESS
0 0 0 0 0
1
1 0 0 0 0
1
0 1 0 0 0
2
1 1 0 0 0
3
0 0 1 0 0
4
1 0 1 0 0
5
Summary of Contents for REMOTE 1100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 1100 1104 Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 1100 1104 Operating Manual t iv 248083363 1 Rev 4...
Page 12: ...Lighthouse REMOTE 1100 1104 Operating Manual ii 248083363 1 Rev 4...
Page 16: ...Lighthouse REMOTE 1100 1104 Operating Manual 1 4 248083363 1 Rev 4...
Page 60: ...Lighthouse REMOTE 1100 1104 Operating Manual A 18 248083363 1 Rev 4...
Page 62: ...Lighthouse REMOTE 1100 1104 Operating Manual B 2 248083363 1 Rev 4...
Page 64: ...Lighthouse REMOTE 1100 1104 Operating Manual C 2 248083363 1 Rev 4...
Page 68: ...Lighthouse REMOTE 1100 1104 Operating Manual I 4 248083363 1 Rev 4...
Page 69: ...00...
Page 70: ......