ApexRemote MODBUS Register Map v1.50
248083447-1 Rev 1
A-15
2
Particle Overflow Status
0 = No overflow 1 = Overflow occurred.
3
Instrument Service Status
0 = Instrument is working correctly.
1 = Service light is on. Instrument malfunction detected.
4
Threshold High Status
0 = Threshold not exceeded. 1 = Threshold exceeded
5
Threshold Low Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
6
Instrument Sampler Status
0 = Nominal Operation. 1 = Sampler Error.
7
ApexRemote:
0 = LASER power in spec. 1 = LASER power out of
spec.
8
ApexRemote:
LASER current status
0 = LASER current in spec. 1 = LASER power out of
spec.
9
ApexRemote:
LASER supply status
0 = LASER supply in spec. 1 = LASER supply out of
spec.
10
ApexRemote:
LASER life status
0 = LASER life in spec. 1 = LASER life out of spec.
11
ApexRemote:
No flow status
0 = There is flow in the unit. 1 = no flow in unit.
12
ApexRemote:
Photoamp supply status
0 = Photoamp supply in spec. 1 = Photoamp supply out
of spec.
13
ApexRemote:
Background status
0 = Photoamp background good. 1 = Photoamp back-
ground out of spec.
Table A-14: Instrument Current Status
Value
Action
Summary of Contents for ApexR02
Page 1: ......
Page 2: ...00 ...
Page 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Page 6: ......
Page 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Page 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Page 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Page 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Page 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Page 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Page 83: ...00 ...
Page 84: ......