Communications
248083447-1 Rev 1
4-7
Ethernet Settings:
The
Apex
Remote
PoE
comes with the
Ethernet Enabled
checkbox
enabled (default) and the
DHCP
disabled. Enter the correct settings in
IP Address
,
Subnet, Gateway
. The IT group should provide these
numbers to prevent conflicts with devices already on the network. All
TCP/IP values should be static. Checking the DHCP checkbox (Figure
4-4) can cause the instrument and its data to become ’lost’ when its IP
is changed during updates from the DHCP server. Ensure DHCP is
OFF and click
Update Instrument
when done.
Figure 4-4
Instrument
Setup
Tool
TCP/IP Set Up Screen
Use IP address provided by IT group
Ethernet checked
DHCP Check OFF
Click to
Save to
Instrument