Program with the
MODBUS Protocol
The
ApexRx
family of instruments can be programmed using the
MODBUS Protocol. The full protocol, as used, is detailed in Appendix
A: “ApexRemote MODBUS Register Map v1.50” on page A-1.
This chapter contains the information needed to program the basic
configuration for the instrument using the MODBUS protocol.
DIP Switches
During power-up and reset, the
Serial ApexRx
reads the DIP switches.
Note:
When changing
the DIP switch settings, the
instrument power must be
power-cycled.
Protocol
Settings
The MODBUS Protocol is defined through an RS-232 or RS-485
interface with:
•
Baud Rate:
19200
•
Data Bits:
8
•
Stop Bits:
1
•
Parity:
None
•
Flow Control: None
Power On/
Auto Start
When powering up the instrument, it will begin sampling using the
default configuration:
•
Sample Time =
60 seconds
•
Hold Time =
0 seconds
•
Alarm Channel =
Disabled
Note:
The automatic
starting of the sampling
accommodates systems
that do not send a START
command, but just polls
the instrument for its data.
To stop the sampling, send the command
12
to command register
40002.
Stopping the sampling will set the Device Status bit in Register 40003
to 0.
Summary of Contents for ApexR02
Page 1: ......
Page 2: ...00 ...
Page 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Page 6: ......
Page 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Page 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Page 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Page 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Page 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Page 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Page 83: ...00 ...
Page 84: ......