ApexRemote MODBUS Register Map v1.50
248083447-1 Rev 1
A-7
Additional status bits are shown in 40056 displayed in Table A-6.
2
NEW DATA:
Set to 1 to indicate that a new data record
has been recorded and it hasn't been read via modbus yet.
When a data record has been read via modbus (registers
30001 to 30999), then this flag is reset to zero.
3
DEVICE ERROR:
in the event that there is a failure on
the device, this bit is set to indicate possible invalid data
has been collected.
9
ApexRemote:
DATA VALIDATION:
Set to 1 when unit is in data
validation mode, else set to 0.
10
ApexRemote:
LOCATION VALIDATION:
Set to 1 when unit is in
Location Validation mode, else set to 0.
11
ApexRemote:
LASER STATUS:
Set to 1 when unit’s LASER is out of
spec, else set to 0.
12
ApexRemote:
FLOW STATUS:
Set to 1 when unit’s flow is out of
spec, else set to 0.
13
ApexRemote:
SERVICE STATUS:
Set to 1 when unit needs to be
serviced, else set to 0.
14
ApexRemote:
THRESHOLD HIGH STATUS:
Set to 1 when unit’s
high alarm threshold is exceeded, else set to 0.
15
ApexRemote:
THRESHOLD LOW STATUS:
Set to 1 when unit’s
low alarm threshold is not met, else set to 0.
Table A-6 Device Status Status (40056)
Bit
Description
0
ApexRemote:
LASER POWER STATUS
: Set to 1 when unit’s
LASER current is out of spec, else set to 0
Table A-5 Device Status Status (40003/40057)
Bit
Description
Summary of Contents for ApexR02
Page 1: ......
Page 2: ...00 ...
Page 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Page 6: ......
Page 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Page 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Page 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Page 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Page 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Page 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Page 83: ...00 ...
Page 84: ......