background image

2

2

21

To proceed with the simulated blood 

pressure measurement, first close the 

valve on the sphygmomanometer bulb 

tightly and begin pumping air into the 

cuff until the gauge reads higher than 

the preset systolic level chosen. Once 

this point is reached, loosen the valve 

on the bulb slightly to allow the gauge 

pressure reading to decrease slowly. 

While monitoring the arm assembly 

with the stethoscope, note the point 

on the sphygmomanometer gauge 

when the first Korotkoff sound is 

heard. This will be the systolic blood 

pressure. Allow the pressure in the 

cuff to continue to decrease until the 

point at which the last pulse is heard, 

noting the reading on the gauge. This 

is the diastolic blood pressure.
Compare the results of reading the 

systolic and diastolic blood pressures 

on the gauge with the respective set-

tings on the simulator. If the readings 

were accurate they should compare 

favorably with the preset values. If the 

auscultatory gap had been selected, 

then an absence of an audible pulse 

would have been noticed during what 

would have been the phase 2 Korotkoff 

sound. It is this lack of an audible 

pulse that is considered an ausculta-

tory gap. Note that in reality  

the auscultatory gap can be present 

in either the phase 1 or phase 2 

Korotkoff sounds. If it is desired to 

demonstrate the sounds heard while 

measuring a subject’s blood pressure 

to the trainee or group of trainees, an 

auxiliary blood pressure speaker ampli-

fier system is available (SB20146U). 

If the auxiliary speaker is used, the 

speaker is plugged into the EXT 

AUDIO AMP jack located next to the 

ARM output jack at the top of the unit. 

Adjust the volume control to increase 

the output of the auxiliary speaker 

amplifier, and proceed with the blood 

pressure measurement sequence as it 

would normally be performed except 

that the stethoscope need not be used. 

Instead, listen to the sounds as they 

emanate from the speaker amplifier, 

noting the differences in the Korotkoff 

phases being presented.

The Nasco

 

Life

/

form

® 

Blood Pressure 

Simulator is programmed to demon-

strate the 5 Korotkoff phases, includ-

ing an auscultatory gap, which can 

be heard during auscultation of a 

subject, while measuring the subject’s 

blood pressure. Each is distinctly dif-

ferent and present for only a portion 

of the measurement sequence.

Low Battery Indicator

When the battery supply diminishes to 

a level near the point that the unit will 

no longer function properly, the “low 

batt” segment of the systolic pressure 

display will activate when the pressure 

in the sphygmomanometer cuff reach-

es above 20 mmHg. At this point, the 

batteries should be replaced as soon 

as possible in order to insure proper 

operation of the unit. Refer to the sec-

tion titled “Installing the Batteries.”

Calibration Procedures

To calibrate the simulator, set the unit 

up as described in the section titled 

“Using The Nasco

 

Life

/

form

® 

Blood 

Pressure Simulator.” Apply the cuff 

to the simulated arm. Set the control 

box systolic pressure to 150 mmHg 

and set the diastolic pressure to 70 

mmHg. Proceed with a simulated 

blood pressure measurement. Note 

the discrepancy in the readings 

between the gauge and the control 

box. Set the systolic “offset.” 

Example:

 

If the blood pressure reading was 

taken and the sounds started at 148 

mmHg, then the offset is +2. If the 

sounds started at 152 mmHg, the 

offset is –2. For this example, assume 

that the sounds started at 148 mmHg. 

Press and hold the Calibration key 

until the systolic correction window 

appears. 

(See figure 27.)

 Using the 

arrow up key set the correction to +2. 

Set the diastolic offset. 

Example:

 If 

the blood pressure reading was taken 

and the sounds stopped at 72 mmHg, 

the offset is –2; or if the sounds 

stopped at 68 mmHg, the offset is +2. 

Procedure A

Summary of Contents for CRiSis LF03953U

Page 1: ...plete CRiSis LF03955U Deluxe CRiSis LF03956U Deluxe Plus CRiSis LF03965U CRiSis Auscultation LF03966U CRiSis Auscultation w Interactive ECG LF03968U Deluxe Plus CRiSis Auscultation INSTRUCTION MANUAL...

Page 2: ...defects in material and workmanship for a period of three years from the date of purchase Nasco will repair any defect reported within three years of the date of purchase at no charge Products found t...

Page 3: ...ult 27 Defibrillation Training Pediatric 28 External Pacer Training Pediatric 28 Procedure C Auscultation Introduction 29 Instructions for Use 29 Supplies Replacement Parts for Auscultation 29 Introdu...

Page 4: ...MODEL Pages 5 10 display the Adult CRiSis family of simulators Identify your Life form Adult CRiSis Manikins model and refer to the Operation Section pertaining to your model This guide has been care...

Page 5: ...m Blood Pressure Arm Electronic Blood Pressure Control Unit Defibrillation Chest Skin Pump Spray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small T...

Page 6: ...e Arm Electronic Blood Pressure Control Unit Defibrillation Chest Skin Pump Spray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small Towels Butterfly...

Page 7: ...lation Chest Skin Pump Spray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small Towels Butterfly Infusion Set Synthetic Blood Sphygmomanometer Batter...

Page 8: ...e Control Unit Defibrillation Chest Skin Pump Spray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small Towels Butterfly Infusion Set Synthetic Blood...

Page 9: ...re Control Unit Defibrillation Chest Skin Pump Spray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small Towels Butterfly Infusion Set Synthetic Blood...

Page 10: ...pray Lubricant 3 cc Syringe with Needle 12 cc Syringe with Needle Two IV Bags Three Pinch Clamps Two Small Towels Butterfly Infusion Set Synthetic Blood Sphygmomanometer Batteries Included Electronic...

Page 11: ...se see Using the Combitube Suction techniques and proper cuff inflation can also be per formed and evaluated Lubricating the Airway Trainer Head Lubricate both the simulator and supplies being used wi...

Page 12: ...the upper compression plate See figure 5 Figure 5 4 Place the lung bags through the hole in the center of the upper compression plate so they rest on top of the plate See figure 6 Figure 6 5 Place th...

Page 13: ...disinfect repeat the standard cleaning procedure but this time use a bleach solution as specified by the Centers for Disease Control instead of soapy water Fill the system with the solution until it...

Page 14: ...oner The Skin and Vein Kits are available through Nasco see page 9 for list of supplies Internal Structure Internally the vascular structure rub ber tubing begins at the shoulder and continues under t...

Page 15: ...be aspirated once the vein is prop erly punctured 8 Small diameter needles 20 to 25 gauge should be used Figure 14 B Preparing the Arm for Intravenous Infusions 1 Close the clamp at the end of IV bag...

Page 16: ...desirable rate with clamp With this arrangement the IV bag B when full may be easily switched with A NOTE always regulate flow of blood from the raised bag and open the other clamp D Intramuscular In...

Page 17: ...ulated to be compatible with the self sealing veins and plastics used in manufacturing the arm 2 NEVER use synthetic blood for intramuscular injection 3 DO NOT use dull or burred needles as these will...

Page 18: ...e After the bat teries have been properly installed reassemble the Blood Pressure Simulator by simply reversing the dis assembly procedures Place the unit face up on the work surface and turn it on by...

Page 19: ...ng Yourself with the Nasco Life form Blood Pressure Simulator Control Panel Under the display window are three buttons Menu Gap and Calibrate See figure 20 The systolic pres sure is set by pressing th...

Page 20: ...he trainee can become familiar with this phenomenon Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabl...

Page 21: ...e as it would normally be performed except that the stethoscope need not be used Instead listen to the sounds as they emanate from the speaker amplifier noting the differences in the Korotkoff phases...

Page 22: ...key to change the diastolic window See figure 28 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window will be...

Page 23: ...ctors it will be necessary to purchase the corresponding adapt ers from Nasco To order please see Supplies Replacement Parts on page 15 Figure 33 Nasco Life form Defibrillation Chest Skin About the Si...

Page 24: ...connections on the patient simulators one or more may be disconnected 2 Check to make sure your patient simulator is plugged in and working properly Problem ECG wave is inverted Solution Recheck the p...

Page 25: ...utomatically from one rhythm running rhythm to another rhythm waiting rhythm when a defib discharge is sensed If defibrillating into Nasco manikin set defib energy to 2J or more If defibrillating dire...

Page 26: ...nitor defibrillator Observe that NSR at 72 bpm is displayed Using Simulator without Defib Manikin 1 Connect monitor s ECG cable to simulator s ECG snaps See Figure 1 2 Using your pad adapters or pad a...

Page 27: ...VT alternates with NSR then remains in SVT SVT rate 216 S TACH Sinus Tachycardia Rate 120 NSR Normal Sinus Rhythm Rate 72 ASYS Asystole SINUS PVC Sinus Rhythm with PVCs Sinus rate 72 FOR EXTERNAL PACE...

Page 28: ...er 2 1 Ventricular rate 150 SVT Supraventricular Tachycardia Inverted P follows QRS Rate 240 S TACH Sinus Tachycardia Rate 165 NSR Normal Sinus Rhythm Rate 90 ASYS Asystole SINUS PVC Sinus Rhythm with...

Page 29: ...he scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenos...

Page 30: ...l is displayed separately while averages are stored in the memory Designed to automatically switch between the compression and ven tilation modes depending on which one is being applied Electronic Mon...

Page 31: ...every time the compres sion depth of 1 00 inch is reached It will turn off when it drops below this depth or if it exceeds 1 50 inches TROUBLE SHOOTING The red placement light is continu ously on ind...

Page 32: ...on is pushed Upon pushing the stop button the unit will calculate and display the average depth of all com pressions from the start of the present cycle Any compressions made after the stop button is...

Page 33: ...duration in sec onds of the last ventilation The unit will continue this display mode until the mode is changed or the stop but ton is pushed Upon pushing the stop button the unit will calculate and d...

Page 34: ...t shows 1 Time scale 1 8 1 second 2 Axis for compression waveforms start at the center and goes left 3 Axis for ventilation waveforms start at the center and goes right 4 1 0 limit line for compressio...

Page 35: ...towards the front of the printer See figure 10 5 While keeping paper taut position the paper within the guides of the printer and re install the platen so that the gears of the platen engage with the...

Page 36: ...36 Procedure F Troubleshooting Information is on pages 31 33 Supplies Replacement Parts for Printer LF03451U Thermal Printer Paper...

Page 37: ...lus CRiSis Auscultation Manikin LF03968U CRiSis Torso s No Heart and Lung Sounds CRiSis Torso Manikin LF03957U Deluxe CRiSis Torso Manikin LF03958U Nasco CRiSis Manikins E l e c t r o n i c M o n i t...

Page 38: ...PR Dog LF01162U Venatech IV Trainer LF01174U NG Tube Trach Skills LF01184U Venatech IM Sub Q LF01193U Special Needs Baby LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer with Stand LF...

Reviews: