GETTING STARTED
Using Simulator with Defib Manikin
1. Connect manikin’s ECG cable to simulator’s color-coded ECG snaps. (
See
Figure 1.)
This is the only cable needed for connecting simulator to manikin.
2. Connect defibrillation cable to manikin’s defibrillation sites. (If you have a
separate pacer cable, attach it to defibrillation sites for pacing.)
3. Connect monitor’s ECG cable to manikin’s ECG snaps. If monitor ECG cable
has right leg lead, but manikin does not have
right leg ECG snap, connect right leg lead to
simulator’s green-labeled ECG snap.
4. Press
ON-OFF
key to power-on simulator.
Observe that
NSR
and
Adult
indicators are
illuminated. Power-on monitor/defibrillator.
Observe that NSR at 72 bpm is displayed.
Using Simulator without Defib Maniki
n
1. Connect monitor’s ECG cable to simulator’s
ECG snaps.
(See Figure 1.)
2. Using your pad adapters or pad adapter
cable, connect cable to simulator’s defib
cable receptacles.
(See Figure 2.)
3. Press
ON-OFF
key to power-on simulator.
Observe that
NSR
and
Adult
indicators are
illuminated. Power-on monitor/defibrillator.
Observe that NSR at 72 bpm is displayed.
Figure 2
Insert pad adapters, or pad
adapter cable, into adapter
receptacles. Connect defibril-
lation cable — APEX to left,
STERNUM to right.
WARNING: SHOCK HAZARD!
Be sure defibrillation cable is
securely attached to simulator.
Figure 1
Press ECG cable connectors
onto simulator snaps (match
colors). Left to right: green
(RL), white (RA), black (LA),
red (LL).
ECG signal is obtained with
either 3- or 4-lead ECG cable.
Procedure B