Figure 29
Lastly, take the loose end of the tub-
ing and push the female luer fitting
into it.
(See figure 32.)
Attach the
female luer fitting to the fitting marked
CUFF at the top of the unit. Once all
of the necessary connections have
been properly made, go to the section
titled “Calibration Procedures” and
calibrate the unit. When the calibration
procedures have been completed,
the simulator is ready for use.
Supplies/Replacement Parts for
Blood Pressure Simulato
r
LF01096U
Electronic Control Unit with
Sphygmomanometer
SB20146U
Blood Pressure Speaker
System
Figure 30
Figure 31
Figure 32
For this example, assume that the
sounds stopped at 72 mmHg. From
the systolic window, press the Menu
key to change the diastolic window.
(See figure 28.)
With the arrow down
key, set the correction to –2. Press the
Menu key again and the message
“CALIBRATION COMPLETE” will
appear. The main menu window will
be displayed.
Preparing Your Equipment for
Use with the Nasco
Blood Pressure Simulator
To adapt your sphygmomanometer
for use with the simulator, it is first
necessary to obtain the luer fitting and
the T-fitting included with the simula-
tor, which is also available through
your Nasco catalog sales office.
Using a wire cutting pliers or similar
instrument, carefully cut the pressure
line of the sphygmomanometer about
2" from the gauge.
(See figure 29.)
Take the T-fitting and insert it between
the two ends of tubing that were previ-
ously cut.
(See figure 30.)
Assemble
the portion of pressure line tubing
over the remaining barbed end on the
T-fitting.
(See figure 31.)
Figure 27
Figure 28
Procedure A