9. Remove the Capsule Holding Pins' screws with a crosshead screwdriver.
10. Remove the Capsule Holding Pins from the Capsule Magazine.
9
10
Replace the Capsule Holding Pins
11. Insert one of the new Capsule Holding Pins into the Capsule Magazine.
11.1 Note: Ensure that the new Capsule Holding Pin is straight and inserted at the
correct length.
12. Tighten the new Capsule Holding Pin's screw with a crosshead screw driver.
13. Repeat steps 11-12 until all new Capsule Holding Pins are installed.
14. Reinsert the Capsule Magazine's teeth and secure them with an Allen key.
14.1 Note: Ensure that the spring correctly inserted as well.
15. Ensure that the new Capsule Holding Pins are all aligned and in the correct position.
15.1 Note: If the alignment is o
ff
, take o
ff
the Capsule Magazine's teeth and adjust
accordingly.