50
Installation
2.5
Connecting the cooling water
Cooling water specifications
Inlet temperature
10 - 30
°
C
Inlet pressure
2 to 7 bar absolute
Cooling water requirement
See Fig. 38
Appearance
Colorless, clear, free of oils and greases
Sediments
< 250 mg/l
Particle size
< 150 µm
Electrical conductivity
< 500 µS/cm
pH value
7 to 8.5
Overall hardness (total alkaline earths)
max. 20
°
German hardness scale
(= 3.57 mmol/l)
Further information on request.
Connect the cooling water to the connectors; see Fig. 37.
The CT versions have a cooling water bypass.
Make sure that you
do not mix up
the inlet and the outlet con-
nection.
Turn off the cooling water supply when the pump is switched off in order to
avoid condensate formation in the pump.
If you do not close the cooling water it may take longer to achieve ultimate
pressure after start up of the system.
No cooling water is reqired when operating the pump at ultimate pressure.
Cooling water is only reqired when baking out the pump.
Pump cooling
Bypass
Fig. 36 Schematic of the cooling water flow for MAG (W) 1500, 2800, 3200 CT
Caution for CT versions