GA05141_0702 - 09/2004
79
Plug-in control
Fig. 56 Diagram: Operation status
Mains switched on
Initialization
Switch On Guard
Acceleration
Normal Operation
Mains Down
Overload
Deceleration
Failure
After a failure, quit a second time
START command
Mains ok
Frequency > P 25 x P24
Mains failure
Mains failure
> 5 s
Frequency > P25 x P24
Frequency < P25 x P24
Mains failure
STOP command
If a failure occurs
the pump is decelerated
Frequency < 5 Hz
Failure occurs
Failure acknowledgement
only possible if f < 5 Hz
Ready
Start Delay
STOP command
Parameter:
P24 = Frequency setpoint
P25 = Factor normal operation
Stand-by
Option
command
Summary of Contents for MAG 1500 CT
Page 107: ...GA05141_0702 09 2004 107 Notes...