GA05141_0702 - 09/2004
35
Installation
Fig. 21 Connection elements, installing the splinter guard and fixing the intake flange for the MAG W 1300 C
Forevacuum flange
Purge gas in
High-vacuum flange
DRIVE/BEARING
connection
PK
connection
Cooling water
connection
MAG W 1300 C
DN 200: 260 mm
DN 250: 310 mm
DN 200: 213 mm
DN 250: 261 mm
14 mm
35 mm
Vacuum sealing disk consisting
of centering ring and O-ring
with outer support ring
Retaining ring
ISO-K flange
ISO-F collar flange
Bolt
14 mm
35 mm
Correct
Wrong
Caution
Install the splinter guard as shown.
Installing the splinter guard upside
down may lead to contact between
splinter guard and rotor during fast
venting of the pump.
12 bolts M10 x 35
Installation torque per bolt 35
+5
Nm
Bolt quality:12.9 according to EN ISO 898-1
with coating
0,2% yield strength > 1080 N/mm
2
ISO-F flange
ISO-K flange
Summary of Contents for MAG 1500 CT
Page 107: ...GA05141_0702 09 2004 107 Notes...