GA05141_0702 - 09/2004
16
Description
Technical data (continued)
MAG
1500 CT
W 1500 CT
1500 CT
W 1500 CT
High-vacuum flange
DN
200 ISO-F
200 ISO-F
250 ISO-F
250 ISO-F
200 CF
Pumping speed (PNEUROP)
for N
2
l·s
-1
1100
1100
1220
1220
for Ar
l·s
-1
1000
1050
1180
1180
for H
2
l·s
-1
920
920
1020
1020
Compression for N
2
>10
8
>10
8
>10
8
>10
8
Ultimate pressure as to DIN 28 400
mbar
< 10
-8
< 10
-8
< 10
-8
< 10
-8
Max. forevacuum pressure with N
2
mbar
1.7
2.6
1.7
2.6
Rotor speed
min
-1
36,000
Run-up time
min
< 6
Braking time with/without venting
min
1 / < 6
Cooling
water
Cooling connection, threads
G
1/8“
with Swagelok elbow for tube OD
mm
6,4 (1/4“)
Cooling water temperature
°
C
10 – 30
°
F
50 – 86
Base flange temperature
°
C
20 – 90
°
F
68 – 194
Weight
approx. kg
32
Max. high-vacuum flange temperature
short-time
°
C (
°
F)
85 (185)
for continuous operation
°
C (
°
F)
60 (140)
for bake-out (CF flange only)
°
C (
°
F)
–
120 (248)
–
–
Vibration level at high-vacuum flange
at max. speed
µm
< 0.01
Forevacuum flange
DN
40 KF
Recommended backing pump
Dry commpressing pump with
pumping speed
m
3
/h
100
or rotary vane pump
TRIVAC
D 65 BCS
Admissible ambient temperature
°
C
5 – 40
°
F
40 – 104
Storage temperature
°
C
-10 – +60
°
F
14 – 140
Max. relative air humidity
95% (non-condensing)
Degree of protection (EN 60529)
IP 20
Summary of Contents for MAG 1500 CT