![Leuze electronic LSIS 472i M45-I1 Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/leuze-electronic/lsis-472i-m45-i1/lsis-472i-m45-i1_operating-instructions-manual_1884019033.webp)
Extended configuration – Leuze electronic webConfig tool
Leuze electronic
LSIS 472i
33
Select
CONFIGURATION > PROGRAMS
.
Click on the [Configure compartment fine positioning] button.
Figure 9.4:
Configure compartment fine positioning
dialog
Enter the type and diameter of the marking.
Enter the distance between rack profile and lens cover of the device.
Enter the positioning accuracy for the active check program.
The positioning accuracy is determined over a rectangular positioning area depending on the diameter
of the marking. If the device sees the marking within this area, positioning of the high-bay storage device
is concluded (I05 … IO8 = 1).
Example:
• Diameter of the marking: 15 mm
Positioning accuracy X: 5 mm
Positioning accuracy Y: 5 mm
If the device sees the marking within an area of 25 mm x 25 mm around the center of the marking, all
switching outputs switch to 1.
Enter the offset values for the active check program.
The offset can be used to enter a deviation of the positioning relative to the middle of the marking, e.g.,
to take into account the difference in positioning between loading position and unloading position of the
high-bay storage device.
Confirm all entries with [OK].
9.3.5 Calibrating
During device calibration, the measurement distance is calibrated with the diameter of the marking.
Select
CONFIGURATION > PROGRAMS
.
Select the active check program (
Rack Near
or
Rack Far
).
Select
CONFIGURATION > PROGRAM > Tools > Calibration
.
The
Tool configuration – Calibration
dialog is displayed.
For a single-depth warehouse, enter the same distance under
Rack Near
and
Rack Far
.